The Analyst · 2018 · 20 citations · 39 references
EngineeringBiochemistryMolecular BiologyBiosensorsElectroanalytical MethodsEpigeneticsCell-based BiosensorsDna MethylationElectrochemical SensorsNanomaterial-based BiosensorsChemiluminescent Signal AmplificationDna MethyltransferasesElectrochemical PotentialsElectrochemical Signal AmplificationElectrochemiluminescent BiosensorsLuminescence-based AssaysLuminescent Reporter SystemsSignal-to-noise Ratio
In this paper, a dual-signal ratiometric electrochemiluminescence (ECL) assay was developed to detect human DNA (cytosine-5)-methyltransferase1 (DNMT1) activity. Mercaptopropionic acid (MPA)-modified Eu3+-doped CdS nanocrystals (MPA-CdS:Eu NCs) and luminol were used as two different ECL emitters, which showed ECL signals at different electrochemical potentials. The resultant ECL peaks exhibited opposite trends in the DNMT1 system. According to the ratio of the ECL responses of CdS to luminol, the DNMT1 activity was detected in the range of 1.0-30.0 U mL-1 in buffer solution. The limit of detection (LOD) was estimated to be 0.07 U mL-1 at a signal-to-noise ratio of 3. The assay was extended to detect the DNMT1 activity from crude lysates of cancer cells, along with the effect of DNMT1 inhibitors such as 5-aza and 5-aza-2'-dC.
39
Electrochemiluminescence (ECL)
Mark Richter · Chemical Reviews · 2004 · 2.5K citations
The Role of DNA Methylation in Mammalian Epigenetics
Peter A. Jones, Daiya Takai · Science · 2001 · 2.1K citations
Recombinant Human DNA (Cytosine-5) Methyltransferase
Sriharsa Pradhan, Albino Bacolla, Robert D. Wells et al. · Journal of Biological Chemistry · 1999 · 640 citations · Full text