Magnetic Resonance in Medicine · 2018 · 21 citations · 37 references
MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyBrain ImagingDigital Signal ProcessingImage ReconstructionMagnetic Resonance Image ReconstructionMagnetic Resonance ImagingMedical ImagingLow-rank ReconstructionPhase CorrectionSensitivity CalibrationSimultaneous Multi-slice ReconstructionVirtual Channels
PURPOSE: To improve simultaneous multislice (SMS) EPI by robust Nyquist ghost correction in both coil sensitivity calibration and SMS reconstruction. METHODS: To derive coil sensitivity and slice-dependent phase difference map between positive- and negative-echo images, single-band EPI reference data are fully sampled with EPI parameters matched to SMS acquisition. First, the reference data are organized into positive- and negative-echo virtual channels where missing data are estimated using low-rank-based simultaneous autocalibrating and k-space estimation (SAKE) at small matrix size. The resulting ghost-free positive- and negative-echo images are combined to generate coil sensitivity maps. Second, full-matrix positive- and negative-echo images are SENSE reconstructed from the reference data. Their phase difference or error map is then calculated. Last, SMS EPI is reconstructed using phase error correction SENSE (PEC-SENSE) that incorporates phase error map into coil sensitivity maps for negative-echo data. The proposed method was evaluated using both experimental data from 7T systems and simulations. RESULTS: Virtual coil SAKE eliminated Nyquist ghosts in the single-band EPI, yielding high-quality coil sensitivity maps and phase error maps. The subsequent PEC-SENSE robustly reconstructed SMS EPI under various conditions, including presence of in-plane acceleration, with lesser artifacts and higher temporal SNR than slice-dependent 1D linear correction method. CONCLUSION: The proposed procedure of virtual coil SAKE calibration and PEC-SENSE reconstruction substantially reduces all ghost-related artifacts originating either directly from SMS EPI data or indirectly from EPI-based coil sensitivity maps. It is computationally efficient, and generally applicable to all SMS EPI-based applications.
37
SENSE: Sensitivity encoding for fast MRI
Klaas P. Pruessmann, Markus Weiger, Markus B. Scheidegger et al. · Magnetic Resonance in Medicine · 1999 · 6.1K citations
Generalized autocalibrating partially parallel acquisitions (GRAPPA)
Mark A. Griswold, Peter M. Jakob, Robin M. Heidemann et al. · Magnetic Resonance in Medicine · 2002 · 5.3K citations · Full text
Kawin Setsompop, Borjan Gagoski, Jon̈athan R. Polimeni et al. · Magnetic Resonance in Medicine · 2011 · 1.4K citations · Full text