This paper studies the design and application of an interferometric sensor system intended for measuring intraocular pressure. The system consists of a hermetically sealed, biocompatible, wireless, and powerless interferometric pressure sensor and a portable handheld reader. The interferometric pressure sensor consists of a 200 nm silicon nitride diaphragm, a $10~\mu \text{m}$ SU-8 spacer, and a $200~\mu \text{m}$ glass substrate. As monochromatic light is directed towards the sensor cavity, interference fringes between the silicon nitride diaphragm and the glass substrate form as a function of the applied pressure. In-vitro measurements show pressure in the range of 0-60 mmHg can be determined reliably with an accuracy of ±0.3 mmHg, while ex-vivo experiments using rabbit cornea showed an accuracy of ±0.6 mmHg. The resolution of the sensor is 0.2 mmHg. These results demonstrate the ability of the system to measure small changes in pressure, including intraocular pressure of glaucoma patients.