Concepedia

Publication | Open Access

Clinical impact, safety, and accuracy of the remotely monitored implantable loop recorder Medtronic Reveal LINQTM

65

Citations

16

References

2017

Year

Abstract

The remote monitoring feature of the Reveal LINQTM allowed earlier diagnosis of asymptomatic but serious arrhythmias in a significant proportion of patients. Implantation of the device outside the EP laboratory appeared safe. However, R-wave undersensing and consequent false recognition of bradyarrhythmias remains a clinically important technical issue.