Biomedical Optics Express · 2017 · 18 citations · 22 references
Natural SciencesBiomedical EngineeringPhysicsAcousticsBiomedical OpticsOpticsAcoustic ImagingAcoustic SensingAcoustic TomographyPhotoacoustic ImagingAcoustic Lens DesignAcoustic LensingCompound Lens SystemsDeformationDetectorsDirectivityLensesPhotoacoustic Imaging Systems
In photoacoustic computed tomography, the limited directivity of the detectors may cause deformation of off-center targets and lead to an imbalanced resolution in the imaging area. To improve the directivity of the acoustic detectors, several negative acoustic lenses have been proposed. In this study, we develop a new compound acoustic lens fabricated by integrating a concave polydimethylsiloxane (PDMS) lens and a convex epoxy lens. Both theoretical simulations and experimental evaluations demonstrate that the compound lens provides a larger directivity compared to single lenses made of PDMS, epoxy, and liquid. The measured acceptance angles of a 6-mm piezoelectric acoustic transducer equipped with the compound, epoxy, liquid, and PDMS lenses are 55°, 36°, 25°, and 20°, respectively. No deformation is observed in the off-center targets by using compound lens. However, serious deformation appears in the cases using single lenses.
22
Biomedical photoacoustic imaging
Paul C. Beard · Interface Focus · 2011 · 2.2K citations
Multiscale photoacoustic microscopy and computed tomography
Lihong V. Wang · Nature Photonics · 2009 · 1.4K citations
Laser optoacoustic imaging system for detection of breast cancer
Sergey A. Ermilov, Tuenchit Khamapirad, André Conjusteau et al. · Journal of Biomedical Optics · 2009 · 491 citations · Full text