EngineeringBioengineeringBiomedical EngineeringAcousticsBiomedical ImagingDiagnostic ImagingAcoustic ImagingAcoustic TomographyOptical ImagingPhotoacoustic ImagingAperture SizeFabry-perot ResonatorsPhotoacoustic Image FormationPhotoacoustic Image ReconstructionPhotoacoustic Imaging InstrumentationPhotoacoustic Imaging SystemsPhotoacoustic Sensor Design
Fabry–Pérot sensors have been used to produce in-vivo photoacoustic images of exquisite quality. However, for ease of construction and interrogation, they are produced in a planar form. Planar arrays suffer from a limited detection aperture, which leads to artifacts in the reconstruction of the initial pressure distribution. Here, an L-shaped detection geometry is described that allows a greater field of view by placing a second planar array orthogonal to the first. This captures data from the deeper lying regions of interest and mitigates the limited view, thus reducing artifacts in the reconstructed initial pressure distribution.
30
Edward Zhang, Jan Laufer, Paul C. Beard · Applied Optics · 2008 · 569 citations
Reconstructions in limited‐view thermoacoustic tomography
Yuan Xu, Lihong V. Wang, Gaik Ambartsoumian et al. · Medical Physics · 2004 · 405 citations · Full text
Photoacoustic angiography of the breast
Robert A. Kruger, Richard B. Lam, Daniel R. Reinecke et al. · Medical Physics · 2010 · 302 citations · Full text