Applied Optics · 2008 · 569 citations · 31 references
EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingUltrasonographyUltrasoundAcoustic ImagingPhotoacoustic ImagingThree-dimensional UltrasoundUltrasonicsUltrasound ImagingUltrasound InstrumentationHigh-resolution ImagingHigh-resolution MappingPhotoacoustic Imaging InstrumentationPhotoacoustic Imaging SystemsSpatial Resolution In Photoacoustic ImagingSpatial Resolution MeasurementThree-dimensional Medical ImagingThree-dimensional Photoacoustic Imaging
The authors present a multiwavelength backward‑mode planar photoacoustic scanner designed for three‑dimensional imaging of soft tissues to several millimetre depths with micrometre‑scale resolution. The scanner combines a tunable optical parametric oscillator delivering 600–1200 nm nanosecond pulses with a Fabry‑Perot polymer‑film ultrasound sensor that maps photoacoustic signals over a 50 mm aperture in 10 µm steps, using two sensor spacers (22 µm and 38 µm) that provide 39 MHz and 22 MHz −3 dB bandwidths. The system achieved a noise‑equivalent pressure of 0.21 kPa, lateral and vertical line‑spread functions of 38 µm and 15 µm, and successfully resolved blood‑vessel and dye phantoms, indicating it could serve as a practical alternative to piezoelectric photoacoustic devices for high‑resolution skin microvasculature imaging.
A multiwavelength backward-mode planar photoacoustic scanner for 3D imaging of soft tissues to depths of several millimeters with a spatial resolution in the tens to hundreds of micrometers range is described. The system comprises a tunable optical parametric oscillator laser system that provides nanosecond laser pulses between 600 and 1200 nm for generating the photoacoustic signals and an optical ultrasound mapping system based upon a Fabry-Perot polymer film sensor for detecting them. The system enables photoacoustic signals to be mapped in 2D over a 50 mm diameter aperture in steps of 10 μm with an optically defined element size of 64 μm. Two sensors were used, one with a 22 μm thick polymer film spacer and the other with a 38 μm thick spacer providing −3 dB acoustic bandwidths of 39 and 22 MHz, respectively. The measured noise equivalent pressure of the 38 μm sensor was 0.21 kPa over a 20 MHz measurement bandwidth. The instrument line-spread function (LSF) was measured as a function of position and the minimum lateral and vertical LSFs found to be 38 and 15 μm, respectively. To demonstrate the ability of the system to provide high-resolution 3D images, a range of absorbing objects were imaged. Among these was a blood vessel phantom that comprised a network of blood filled tubes of diameters ranging from 62 to 300 μm immersed in an optically scattering liquid. In addition, to demonstrate the applicability of the system to spectroscopic imaging, a phantom comprising tubes filled with dyes of different spectral characteristics was imaged at a range of wavelengths. It is considered that this type of instrument may provide a practicable alternative to piezoelectric-based photoacoustic systems for high-resolution structural and functional imaging of the skin microvasculature and other superficial structures.
31
Photoacoustic imaging in biomedicine
Minghua Xu, Lihong V. Wang · Review of Scientific Instruments · 2006 · 2.7K citations · Full text
Jan Laufer, Dave T. Delpy, Clare E. Elwell et al. · Physics in Medicine and Biology · 2006 · 318 citations