EngineeringElectrical EngineeringComputational ElectromagneticsMetrologyMicroelectronicsCalibrationDevice CharacterizationElectromagnetic CompatibilityEquipment CalibrationMicrowave Device SimulationMicrowave MeasurementsAntenna Calibration StandardsCalibration StandardsConnectorsDevice CalibrationS-parameter CharacterizationS-parameter MeasurementStandards CalibrationVector Network Analysis
Accurate calibrations of vector network analyzers require that coaxial connectors are considered in the definition of calibration standards. In this paper the S-parameters of 2.4 mm and 1.85 mm connectors are computed using 3D field simulations. It is found that male pin diameter variations have lower impact on slotless 2.4 mm connectors than on slotted 1.85 mm connectors. Performing calibrations with and without taking into account the true S-parameters of the connectors results in a typical difference |¿S <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">11</sub> | = 0.01...0.02 for female 2.4 mm devices under test.