Journal of the American Society for Mass Spectrometry · 2016 · 26 citations · 24 references
PharmacyPharmaceutical AnalysisAnalytical ChemistryAnalytical ToxicologyMass Spectrometric AnalysisInstrumental AnalysisMass AnalyzersMass SpectrometryPeptide AnalysisPeptide IdentificationVolatile Substance AnalysisBovine Serum AlbuminInstrument Dead VolumeMass Spectrometric Drug AnalysisMass Spectrometric QuantificationNeedle-syringe Delivery SystemsOrbitrap Mass AnalyzersOrbitrap Mass SpectrometryPeptide Mass Spectrometry
First results are reported using a simple, fast, and reproducible matrix-assisted ionization (MAI) sample introduction method that provides substantial improvements relative to previously published MAI methods. The sensitivity of the new MAI methods, which requires no laser, high voltage, or nebulizing gas, is comparable to those reported for MALDI-TOF and n-ESI. High resolution full acquisition mass spectra having low chemical background are acquired from low nanoliters of solution using only a few femtomoles of analyte. The limit-of-detection for angiotensin II is less than 50 amol on an Orbitrap Exactive mass spectrometer. Analysis of peptides, including a bovine serum albumin digest, and drugs, including drugs in urine without a purification step, are reported using a 1 μL zero dead volume syringe in which only the analyte solution wetting the walls of the syringe needle is used in the analysis.
24
Analytical Properties of the Nanoelectrospray Ion Source
Matthias Wilm, Matthias Mann · Analytical Chemistry · 1996 · 1.9K citations