Review of Scientific Instruments · 2016 · 50 citations · 60 references
EngineeringMechanical EngineeringFinite Element AnalysisMicroelectromechanical SystemsVibrationAtomic Force MicroscopyCalibrationMechanical MeasurementModal AnalysisNanomechanicsVibration AnalysisAtomic Force Microscopy CalibrationCantilever MechanicsEigenmode AnalysisFinite Element SimulationInterferometric MeasurementStiffnessTip - Sample InteractionsTip-sample Interactions
A method is presented for calibrating the higher eigenmodes (resonant modes) of atomic force microscopy cantilevers that can be performed prior to any tip-sample interaction. The method leverages recent efforts in accurately calibrating the first eigenmode by providing the higher-mode stiffness as a ratio to the first mode stiffness. A one-time calibration routine must be performed for every cantilever type to determine a power-law relationship between stiffness and frequency, which is then stored for future use on similar cantilevers. Then, future calibrations only require a measurement of the ratio of resonant frequencies and the stiffness of the first mode. This method is verified through stiffness measurements using three independent approaches: interferometric measurement, AC approach-curve calibration, and finite element analysis simulation. Power-law values for calibrating higher-mode stiffnesses are reported for several cantilever models. Once the higher-mode stiffnesses are known, the amplitude of each mode can also be calibrated from the thermal spectrum by application of the equipartition theorem.
60
Calibration of atomic-force microscope tips
Jeffrey L. Hutter, John Bechhoefer · Review of Scientific Instruments · 1993 · 4.2K citations
Calculation of thermal noise in atomic force microscopy
Manfred Jaschke · Nanotechnology · 1995 · 1.6K citations