Physical parameters in high-accuracy spectrophotometry

K.D. Mielenz

Journal of Research of the National Bureau of Standards Section A Physics and Chemistry · 1972 · 20 citations · 6 references

DOIFull text

Open access

Concepts

Abstract

The meas ured apparent tran s mittance TA of a filte r or liquid sample de pends on th e beam geometry in th e spectrophotometer. For foc used li ght incid ent upon th e sampl e, TA is differe nt for syste ms having diffe rent f-numbers , and al so depends on the state of polarization of the light. These effects are eliminated wh e n the incide nt light is collimated; in thi s case TA approac hes the "true" transmittan ce ' T of the sampl e. Both modes of operation suffer from stray light and inte rference effects. The form er may be redu ced significantly by using mirror rath er than lens optics, and the latte r by suitable choice of the monochromator s lit width. A new spectrophotometer based upon the above-me nti oned design principl es is de scribed_ The photometric pre cision of thi s instrument is shot-noise limited, pe rmitting measure ments to better than 10 -4 transmittance units_