Microscopy · 2013 · 12 citations · 5 references
MedicinePhysicsBiomedical ImagingBiomedical OpticsDigital ImagingElectron Beam PhysicsElectron MicroscopyAccuracyDefocusElectron Beam FocusingElectron Beam OpticsFocusImage Sharpness Assessment
An accurate method using image sharpness to determine the best focusing is proposed for ultra-high-voltage electron microscopy. This method maximizes image sharpness for adjusting the focus. Five images with different defocus values are used to calculate the image sharpness. To obtain the best focus value that produces greatest image sharpness, fitting the quasi-Gaussian function to five image sharpness is a suitable alternative. This method, which maximizes image sharpness, gives better accuracy than the wobbler method for the ultra-high-voltage electron microscope. The focusing area can be selected without moving the field of view, because the focusing area can be selected at almost any area in the image.
5
On the accuracy of the Sobel edge detector
Josef Kittler · Image and Vision Computing · 1983 · 337 citations
Automated microscopy for electron tomography
Abraham J. Koster, H. Chen, John W. Sedat et al. · Ultramicroscopy · 1992 · 195 citations
Microscopic tomography with ultra-HVEM and applications
Akio Takaoka, Toshiaki Hasegawa, Kiyokazu Yoshida et al. · Ultramicroscopy · 2007 · 49 citations
Multiple scattering effects of MeV electrons in very thick amorphous specimens
Fang Wang, Haibo Zhang, Meng Cao et al. · Ultramicroscopy · 2010 · 15 citations