Development of Sheath-Flow Probe Electrospray Ionization Mass Spectrometry and Its Application to Real Time Pesticide Analysis
27
Citations
30
References
2013
Year
Environmental ChemistryEngineeringAnalytical InstrumentationPesticide-residue AnalysisSample PreparationEnvironmental EngineeringEnvironmental Analytical ChemistryBioanalysisMass SpectrometryAnalytical ChemistryIon Mobility SpectrometryTime Pesticide AnalysisEnvironmental ToxicologyLiving PlantsMedicineDry SampleReal TimeChromatography
For the real time and direct analysis of chemical constituents from living beings and dry sample, sheath-flow probe electrospray ionization mass spectrometry (SF-PESI-MS) has been newly developed. The components from dry or semidry biological tissues can be extracted using the solvent and picked up by the needle for electrospray. This technique was applied to real-time pesticide analysis of living plants. The results have been validated with that of a well-known system, liquid extraction surface analysis mass spectrometry (LESA-MS). It is demonstrated that SF-PESI-MS can produce reasonable ionization efficiency, which is confirmed by LESA-MS.