Journal of Mass Spectrometry · 2009 · 344 citations · 28 references
PharmacyBiomedical EngineeringPharmaceutical AnalysisBiomemsElectroanalytical MethodsMass Spectrometric AnalysisMicrofluidic DevicesMicrofluidicsSample HandlingControl Parameter SettingDried Blood Spot AnalysisDried Blood Spot SamplingMicroextractionNanoelectrospray IonizationParameter ControlQuantitative Data CollectionSample Preparation Systems
A fully automated liquid extraction-based surface sampling device utilizing an Advion NanoMate chip-based infusion nanoelectrospray ionization system is reported. Analyses were enabled for discrete spot sampling by using the Advanced User Interface of the current commercial control software. This software interface provided the parameter control necessary for the NanoMate robotic pipettor to both form and withdraw a liquid microjunction for sampling from a surface. The system was tested with three types of analytically important sample surface types, viz., spotted sample arrays on a MALDI plate, dried blood spots on paper, and whole-body thin tissue sections from drug dosed mice. The qualitative and quantitative data were consistent with previous studies employing other liquid extraction-based surface sampling techniques.
28
Bioanalytical Method Validation—A Revisit with a Decade of Progress
Vinod P. Shah, Kamal K. Midha, John W. A. Findlay et al. · Pharmaceutical Research · 2000 · 1.6K citations
Neil Spooner, Rakesh Lad, Matt Barfield · Analytical Chemistry · 2009 · 432 citations