A micromethod for analysis of total plasma cholesterol using gas-liquid chromatography.

Joseph MacGee, Terry T. Ishikawa, William Miller, Grace Evans, P M Steiner, Charles J. Glueck

PubMed · 1973 · 33 citations · 8 references

Concepts

Abstract

Abstract An accurate, precise, specific, and practicable micromethod requiring only 50 μl of plasma and a minimum number of manipulations for the analysis of total plasma cholesterol was developed for the gas-liquid chromatograph (GLC). Using peak-height ratio measurements, isopropanol-ethyl acetate extracts of saponified plasma samples containing an internal standard (5 α-cholestane) were chromatographed, and the resulting cholesterol levels were nearly identical to those obtained with an AutoAnalyzer (ferric chloride) colorimetric method; with an overall correlation coefficient of 0.99, and a coefficient of variation of 3.12 per cent. This GLC micromethod was specific and reproducible with only 50 μl of plasma, and should have particular utility in pediatric studies by sharply reducing blood-sample size.

References