IMAPSource Proceedings · 2013 · 16 citations · 8 references
Natural SciencesMaterials EngineeringMaterials ScienceBiocompatibilityBiomaterialsMicroelectromechanical SystemsCoatingsImplant MaterialsImplantable BiomaterialsImplantable Biomedical DevicesImplantable MicrodevicesPolymer Thin FilmsBioceramic CoatingsConformal CoatingImplant CoatingsImplant FabricationLifetimesPackaging For Implantable DevicesPolymeric Implant MaterialsSurface Coatings For Implants
We report on the development of a non-hermetic micropackage technology that employs multiple layers of polydimethylsiloxane (PDMS), epoxy, and parylene-C thin films for conformal coating of chronic biomedical implantable systems with laboratory testing results in saline bath at 85°C, achieving very long micropackaging lifetime (up to ~2–5 years). By combining experiments of material characterization and qualitative analysis of the observations, we explain the measured long lifetimes of the micropackages (projected equivalent of ~2–5 years) using these non-hermetic coating materials and composites.
8
Reliable packaging for parylene-based flexible retinal implant
Jay Han-Chieh Chang, Yang Liu, Dongyang Kang et al. · 2013 · 30 citations
Engineering, Biomedical Engineering, Materials Engineering +17