Evaluation of a portable somatic cell counter in the diagnosis of bubaline subclinical mastitis.

Rogério Salvador, R. L. Soliven, Errol Jay Y. Balagan, Nancy S. Abes, Charito A. Gutierrez, Claro N. Mingala

Thai Journal of Agricultural Science · 2014 · 14 citations · 5 references

Full text

Open access

Concepts

Abstract

The study aimed to evaluate the performance of portable somatic cell counter (PortaSCC) relative to that of the laboratory-based somatic cell counter (Fossomatic) in the diagnosis of bubaline subclinical mastitis. It determined the sensitivity and specificity of PortaSCC with Fossomatic as the reference. The agreement between the results of the two test equipments was also measured using kappa statistics. Eighty milk samples were collected from different farms. Samples were immediately processed for somatic cell count using the PortaSCC. Same samples were brought to the laboratory for the somatic cell counts. All samples were processed in triplicates. The PortaSCC has 94.12% and 87.30% sensitivity and specificity, respectively. A substantial agreement (k=0.70) between the results of the two tests was also observed. These test properties of the PortaSCC and its capability to rapidity to provide results rationalize its utilization as an alternative for the laboratory-based cell counter in evaluating milk samples from herds in remote areas under Philippine field conditions.

References

5