A simple and rapid ultra‐high‐performance liquid chromatography–tandem mass spectrometry method to determine plasma biotin in hemodialysis patients

Shigeaki Yagi, Manabu Nishizawa, Itiro Ando, Shiro Oguma, Emiko Sato, Yutaka Imai, Masako Fujiwara

Biomedical Chromatography · 2015 · 11 citations · 6 references

Abstract

A simple, rapid, and selective method for determination of plasma biotin was developed using ultra-high-performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS). After single-step protein precipitation with methanol, biotin and stable isotope-labeled biotin as an internal standard (IS) were chromatographed on a pentafluorophenyl stationary-phase column (2.1 × 100 mm, 2.7 μm) under isocratic conditions using 10 mm ammonium formate-acetonitrile (93:7, v/v) at a flow rate of 0.6 mL/min. The total chromatographic runtime was 5 min for each injection. Detection was performed in a positive electrospray ionization mode by monitoring selected ion transitions at m/z 245.1/227.0 and 249.1/231.0 for biotin and the IS, respectively. The calibration curve was linear in the range of 0.05-2 ng/mL using 300 μL of plasma. The intra- and inter-day precisions were all <7.1%. The accuracy varied from -0.7 to 8.2%. The developed UHPLC-MS/MS method was successfully applied to determine plasma biotin concentrations in hemodialysis patients. Copyright © 2016 John Wiley & Sons, Ltd.

References

6