We made and characterized an X-ray detector on a 25-μm-thick plastic substrate that is capable of medicalgrade performance. As an indirect conversion flat panel detector, it combined a standard scintillator with an organic photodetector (OPD) layer and oxide thin-film transistor backplane. Using solution-processed organic bulk heterojunction photodiode rather than the usual amorphous silicon, process temperature is reduced to be compatible with plastic film substrates, and a number of costly lithography steps are eliminated, opening the door to lower production costs. With dark currents as low as 1 pA/mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> and sensitivity of 0.2 A/W the OPD also meets functional requirements: the proof-of-concept detector delivers high-resolution, dynamic images at 10 frames/s, and 200 pixels/in using X-ray doses as low as 3 μGy/frame.