Research Explorer (The University of Manchester) · 1999 · 27 citations · 4 references
EngineeringElectrical EngineeringBiomedical ImagingDiagnostic ImagingDigital ImagingComputed TomographyComputed Tomography PhysicsElectrical CharacterizationElectrical MetrologyImage ReconstructionCapacitanceCapacitance MeasurementCapacitance Measurement TechniquesSensor ModelSystem NormalizationTomographic Image QualityTomographic ImagingTomography
Abstract-- An electrical capacitance tomography (ECT) system is calibrated at the upper and lower values of permittivity. The measured capacitances are usually normalised and then used for image reconstruction. The conventional normalisation approach assumes that the distribution of the materials is in parallel and hence the normalised capacitance is a linear function of the measured capacitance. This approach is not appropriate in most situations. In this paper an alternative normalisation approach is derived from a series sensor model. Preliminary experimental results show that the quality of reconstructed images can be improved with the new normalisation approach for the situations tested.
4
Cheng Xie, S.M. Huang, M.S. Beck et al. · IEE Proceedings G Circuits Devices and Systems · 1992 · 503 citations · Full text
Engineering, Biomedical Engineering, Electrical Engineering +17
Measurement of voidage in a fluidized bed using a capacitive sensor
G. Delmon, Roland Faure, B. de Gasquet et al. · Powder Technology · 1996 · 11 citations
Developments in Multiphase Metering
Álvaro Silva Ribeiro · All Days · 1996 · 10 citations