Publication | Closed Access
An implantable multielectrode array with on-chip signal processing
311
Citations
11
References
1986
Year
Medical ElectronicsEngineeringBiomedical EngineeringWireless Implantable DeviceNeurochipSocial SciencesExtracellular Neural BiopotentialsBiophysicsOn-chip CircuitryImplantable SensorActive ProbeImplantable DeviceBioinstrumentationMicroelectronicsNeural InterfaceBiomedical SensorsOn-chip Signal ProcessingNeurophysiologyBioelectronicsElectrophysiologyNeuroscience
This active probe can be used for the long-term recording of extracellular neural biopotentials and as a basis for closed-loop neural prostheses. The probe incorporates on-chip circuitry for amplifying, multiplexing, and buffering neural signals recorded from ten recording electrodes spaced 100-/spl mu/m apart. It requires only three leads and operates from a single 5-V supply. On-chip self-test circuitry for testing electrode impedance levels is provided. The on-chip circuitry is fabricated in a die area of 1.3 mm/SUP 2/ using 6-/spl mu/m LOCOS enhancement-depletion NMOS technology, and dissipates 5 mW of power. The probe is 4.7 mm long and 15 /spl mu/m thick, and has a shank which tapers from 160 /spl mu/m near the base to less than 15 /spl mu/m near the tip.
| Year | Citations | |
|---|---|---|
1957 | 2.1K | |
1970 | 535 | |
1985 | 301 | |
1959 | 300 | |
1975 | 154 | |
1986 | 141 | |
1984 | 47 | |
1980 | 30 | |
1980 | 28 | |
1986 | 16 |