Canadian Journal of Physics · 1959 · 14 citations · 0 references
EngineeringPhysicsMetrologyOpticsPhotonicsInterferometryMeasurement UncertaintyOptical DetectorsOptical InstrumentationOptical MetrologyPhotodetectorsFringe AnalysisInterferometric DetectorsInterferometric InstrumentationInterferometric MetrologyNoise Analysis In InterferometryOptical InterferometersOptical InterferometryOptical Path Difference Measurement
The fundamental limit to the precision of setting on a fringe peak due to photon noise is evaluated for a certain class of interferometric methods in terms of parameters characterizing the source, the interferometer, and the detector. It is shown that changes in path difference of 10 −10 should be detectable with an observing time of 1 second, using only modest equipment. Some of the experimental conditions required to attain this precision are discussed.