FT-Raman Spectroscopy at 1.339 Micrometers

Kelly J. Asselin, Bruce Chase

Applied Spectroscopy · 1994 · 28 citations · 4 references

Concepts

Abstract

The usual laser employed for Fourier transform Raman spectroscopy is a Nd:YAG unit lasing at 1.064 μm. In this work, use of the 1.339-μm lasing emission from Nd:YAG has been demonstrated. The sensitivity of this instrument is comparable to that of conventional FT-Raman instruments, and excellent anti-Stokes spectra can be easily obtained. Operation further into the near-infrared offers additional possibilities for fluorescence minimization. Results are shown for copper phthalocyanine.

References

4