EngineeringElectrical EngineeringMaterials ScienceBiosensorsElectronic MaterialsSensorsBiopotential MeasurementElectrocardiographyElectrode MaterialsNoise MeasurementSensor MaterialsWearable BiosensorsBody Sensor NetworksDry Electrode EegDry ElectrodesElectrode - Skin InterfaceElectrode-skin InterfaceNon-contact SensingSolder PastesWearable Physiological Sensors
Dry and non-contact electrodes are an important part of building user-friendly wearable body sensor networks for physiological (ECG, EEG, EMG) recording. This paper presents the theory and characterization of several inexpensive and non-irritating electrode materials that implement dry and non-contact sensors, including lead-free PCB solder finish, latex, cotton, silver cloth and solder mask. The noise spectra of the electrodes were measured on a live subject to fully mode all the effects of biological and chemical noise processes at the skin- electrode interface. In addition, exemplary ECG physiological measurements are also presented and compared against the signal from reference clinical wet Ag/AgCl electrodes. Results show that PCB finish can function as a general purpose, low-noise dry electrode. Latex and solder mask also work well as insulated electrodes.
22
Skin impedance from 1 Hz to 1 MHz
J. Rosell, J. Colominas, Pere J. Riu et al. · IEEE Transactions on Biomedical Engineering · 1988 · 341 citations
Engineering, Biomedical Engineering, Electrical Engineering +10
Novel dry electrodes for ECG monitoring
Anna Gruetzmann, Stefan Hansen, Jörg Müller · Physiological Measurement · 2007 · 255 citations