Bevelled Dual-Channel Microelectrodes

James M. Anderson, A. L. Kleinhaus, Laura Manuelidis, James W. Prichard

IEEE Transactions on Biomedical Engineering · 1974 · 19 citations · 10 references

Concepts

Abstract

Preparation of bevelled dual-channel glass microelectrodes is described. The bevelled tips overcome some of the problems which complicate intracellular recording. The size of the tip opening is increased, resulting in a lower resistance, which in turn confers greater current-carrying capacity and less noise susceptibility. Bevelling also preserves the sharp point, facilitating tissue penetration and minimizing neuronal trauma. The characteristics of bevelled and unbevelled electrodes filled with 0.5 or 3 M KCI, K Acetate and K <L xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</L> SO <L xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">4</L> were similar. Bevelled electrodes were shown to maintain a near constant resistance during current passage, while unbevelled ones showed changes in resistance that were dependent on current magnitude, polarity and duration. Such changes were caused principally by electro-osmosis.

References

10