2005 · 19 citations · 7 references
EngineeringBioengineeringBiomedical EngineeringBiomemsMicroelectromechanical SystemsProstheticsChip PackagingImplantable Biomedical DevicesImplantable MicrodevicesImplantable MicrosystemsIntegrated Circuit DesignIntegrated Circuit FabricationCapacitorsFabricationFlexible Implantable ElectronicsImplant FabricationMicroelectronic Implant PackagingPackaging For Implantable DevicesPhotolithographyRetinal Prostheses
The authors presented an innovative technology for the fabrication of a biocompatible parylene-based high lead count retinal prosthesis in which a prefabricated stand-alone application-specific integrated circuit (ASIC) is placed directly into the fabrication process of the other system components. The package is fabricated in such a way that the ASIC-to-electrode interconnects are patterned using standard photolithography. The density of interconnects is fully scalable to the limits of lithography. This packaging scheme also enables the simultaneous integration and interconnection of discrete components such as chip capacitors with the rest of the system. Electrical test results verify the efficacy of this cost-effective and high-yield packaging scheme, and pave the way for a monolithic implantable parylene-based intraocular system.
7
Visual perception in a blind subject with a chronic microelectronic retinal prosthesis
Mark S. Humayun, James D. Weiland, Gildo Y Fujii et al. · Vision Research · 2003 · 851 citations
IC-integrated flexible shear-stress sensor skin
Yong Ping Xu, Yu‐Chong Tai, Allen Huang et al. · Journal of Microelectromechanical Systems · 2003 · 101 citations
Parylene etching techniques for microfluidics and biomems
E. Mengi, Yu‐Chong Tai · 2005 · 52 citations · Full text