American Journal of Neuroradiology · 2011 · 50 citations · 6 references
MedicineNeurosurgerySurgeryCerebrovascular NeurosurgeryCerebrovascular SurgeryVascular SurgeryAneurysm RepairAneurysm SurgeryIntracranial Pressure ManagementIntracranial Pressure MonitoringStent PlacementAneurysmal FlowCerebral Aneurysm RuptureFusiform Aneurysm ManagementFusiform Aneurysm RepairIntracranial StentingRuptured Aneurysm ManagementStent-assisted Aneurysm Treatment
The approach for treatment of large and fusiform intracranial aneurysms has evolved from stent-assisted coiling to treatment with flow-diverting stents. The treatment results for these stents are promising; however, early postprocedural aneurysm rupture has been described. The exact cause of rupture is unknown but might be related to intra-aneurysmal flow and pressure changes. We measured intra-aneurysmal pressure before, during, and after placement of a flow-diverting stent by using a dual-sensor guidewire. The pressure inside the aneurysm momentarily decreased during placement but was restored to baseline values within minutes. The flow-diverting stent does not seem to protect the aneurysm from the stress induced by pressure or pressure changes within the lumen.
6
CURATIVE ENDOVASCULAR RECONSTRUCTION OF CEREBRAL ANEURYSMS WITH THE PIPELINE EMBOLIZATION DEVICE
Pedro Lylyk, Carlos Miranda, Rosana Ceratto et al. · Neurosurgery · 2009 · 803 citations
Medicine, Neurosurgery, Surgery +14
Götz Benndorf, Uta Herbon, W.-P. Sollmann et al. · PubMed · 2001 · 149 citations · Full text
Medicine, Neurosurgery, Surgery +15