Concepedia

Abstract

Abstract The practical performances of single and multi‐angle photon correlation spectroscopy (PCS) for resolving bimodal distributions of industrial poly (methyl methacrylate) samples was investigated in a comparative study by several users affiliated to academic and industrial groups and by two suppliers of commercial equipment. The results for the harmonic intensity averaged diameters obtained by cumulants analysis reported by the different laboratories are in agreement, in addition to the results for the normalized second cumulants. The uncertainty on the latter quantities is large, however. For the bimodal samples with two populations with average diameters in a ratio of about 2.5 : 1, not all users were able to resolve the distribution in its components by single‐angle PCS. Some slight improvement was obtained by multi‐angle PCS. Other indirect techniques (polarization intensity differential scattering, static light scattering data and disc photosedimentometer) appeared to be superior for resolving the bimodal distributions.