Journal of Applied Crystallography · 1993 · 24 citations · 0 references
Natural SciencesMaterials EngineeringMaterials ScienceAnalytical ChemistryHeat TransferMaterials CharacterizationThermal Expansion AnalysisThermal Properties Of CeramicsX-ray CrystallographyX-ray Data AnalysisX-ray DiffractionX-ray SpectroscopyHeatersPositional UncertaintyPowder DiffractionPowder Diffraction AnalysisPowder X-ray DiffractionTemperature CalibrationThermal ExpansionX-ray Diffraction Analysis
Methods of minimizing the uncertainties in temperature and position that are inherent in high-temperature X-ray powder diffraction measurements made using strip heaters are discussed, with particular reference to the Anton Paar HTK 10 heating stage. Modification of the strip geometry is recommended to largely eliminate temperature gradients across the sample area and the use of the strip heater as an internal standard is recommended to eliminate positional uncertainties and test temperature calibration. Data on the thermal expansion of Al2O3 are used to evaluate the success of these procedures.