Journal of Analytical Methods in Chemistry · 2013 · 43 citations · 20 references
EngineeringBioengineeringBiotechnologyBiosensorsPesticide ToxicologySensorsChemical SensorsEnzymatic BiosensorsEnzyme-based BiosensorsPesticide ChemistryPesticide Residue AnalysisSensor-based AnalysisBiosensor DevelopmentMethod Detection LimitOperational StabilityOrganophosphate BiosensorsOrganophosphorus PesticidesPaper-based BiosensorsPaper-based SensorsPesticide Detection
A paper-based biosensor was developed for the detection of the degradation products of organophosphorus pesticides. The biosensor quantifies acetylcholine esterase inhibitors in a fast, disposable, cheap, and accurate format. We specifically focused on the use of sugar or protein stabilizer to achieve a biosensor with long shelf-life. The new biosensor detected malathion with a detection limit of 2.5 ppm in 5 min incubation time. The operational stability was confirmed by testing 60 days storage at 4°C when glucose was used as stabilizer.
20
A new and rapid colorimetric determination of acetylcholinesterase activity
George L. Ellman, K. Diane Courtney, Valentino Andres et al. · Biochemical Pharmacology · 1961 · 26.9K citations