A Novel Diamond Microprobe for Neuro-Chemical and -Electrical Recording in Neural Prosthesis
66
Citations
50
References
2009
Year
Medical ElectronicsEngineeringBiomimetic MaterialsBiomaterials DesignBiomedical EngineeringNeurochipSocial SciencesNeural ProsthesisNovel Diamond MicroprobeBiomedical DevicesNeurologyMaterials ScienceImplantable SensorImplantable DeviceNeural InterfaceProstheticsPoly-c ProbeBiomedical SensorsNeuroengineeringElectronic MaterialsMicrofabricationNeurophysiologyBioelectronicsElectrophysiologyNeuroscienceNovel Polycrystalline-diamondBiomaterialsBiocompatible Material-Electrical Recording
This paper describes the design, microfabrication, and testing of a novel polycrystalline-diamond (poly-C)-based microprobe for possible applications in neural prosthesis. The probe utilizes undoped poly-C with a resistivity on the order of 10 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">5</sup> Omega middot cm as a supporting material, which has a Young's modulus in the range of 400-1000 GPa and is biocompatible. Boron-doped poly-C with a resistivity on the order of 10 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">-3</sup> Omega middot cm is used as an electrode material, which provides a chemically stable surface for both chemical and electrical detections in neural studies. The probe has eight poly-C electrode sites with diameters ranging from 2 to 150 mum; the electrode capacitance is approximately 87 muF/cm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> . The measured water potential window of the poly-C electrode spans across negative and positive electrode potentials and typically has a total value of 2.2 V in 1 M KCl. The smallest detectable concentration of norepinephrine (a neurotransmitter) was on the order of 10 nM. The poly-C probe has also been successfully implanted in the auditory cortex area of a guinea pig brain for in vivo neural studies. The recorded signal amplitude was 30-40 muV and had a duration of 1 ms.
| Year | Citations | |
|---|---|---|
Page 1
Page 1