IEEE Transactions on Biomedical Engineering · 2005 · 63 citations · 32 references
EngineeringBiomedical EngineeringDermatologyBiomedical ImagingBiophotonicsDiagnostic ImagingLight-tissue InteractionOptical TomographyReflectance ModelingScattering In TissueTissue Optical PropertiesTissue OpticsDiffuse Reflectance ImagingIllumination SystemsOptical Characterization Of Biological TissueOptical PropertiesReflectanceSpectral Dependence Of Tissue OpticsSynthetic Tissue ModelsTissue Reflectance
Three-dimensional, voxel-based, and wavelength-dependent skin lesion models are developed and simulated using Monte Carlo techniques. The optical geometry of the Nevoscope with trans-illumination is used in the simulations for characterizing the lesion thickness. Based on the correlation analysis between the lesion thickness and the diffuse reflectance, optical wavelengths are selected for multispectral imaging of skin lesions using the Nevoscope. Tissue optical properties reported by various researchers are compiled together to form a voxel library. Tissue models used in the simulations are developed using the voxel library which offers flexibility in updating the optical properties and adding new media types into the models independent of the Monte Carlo simulation code.
32
Probability and Random Processes for Electrical Engineering
Stefan D. Leigh · Technometrics · 1991 · 1.2K citations