IEEE Transactions on Instrumentation and Measurement · 2005 · 27 citations · 6 references
EngineeringComputer EngineeringComputer SystemsHeat TransferInstrumentationSensorsSensor CalibrationSensor IntegrationSensor InterfacingTemperature MeasurementTemperature SensorsThermal SensorsAdaptive ArchitectureAdaptive System ArchitectureOnline CalibrationRelative Humidity MeasurementTemperature MonitoringTemperature SensingThermal Environment AssessmentThermal Environment Monitoring
A temperature and relative humidity (RH) monitor using a thermistor and a polyimide-film RH sensor is developed for global assessment of thermal environments. The smart transducer interface module includes the relaxation oscillators for signal conditioning and a one-chip 16-bit microcomputer for networking. The microcomputer accommodates the calibration tables as well as the mandatory transducer electronic data sheets specified by the IEEE 1451.2 standard. The online calibration using the calibration table and the ratiometric signal conditioning allow /spl plusmn/0.14/spl deg/C accuracy over the temperature range from -20/spl deg/C to 50/spl deg/C and /spl plusmn/2.5RH accuracy over the RH range from 20%RH to 90%RH. Beside these high-accuracy measurements using low-cost sensors, the monitor features an adaptive architecture for global networking.
6
IEEE 1451: A standard in support of smart transducer networking
Kang Lee · 2002 · 220 citations
Engineering, Electrical Engineering, Systems Engineering +17
M.A. Stegawski, R. Schaumann · IEEE Transactions on Instrumentation and Measurement · 1998 · 31 citations