Journal of Orthopaedic Research® · 1996 · 105 citations · 23 references
Natural SciencesBiomedical EngineeringMaterials ScienceBiocompatibilityBiomaterialsOrthopaedic BiomaterialsCoatingsImplant-associated InfectionImplant-related Infection PreventionImplantable BiomaterialsProtein AdsorptionSurface Modification Of ImplantsAdherenceAntibacterial CoatingsAntibacterial MaterialsBacterial Adhesion To Implant SurfacesBovine Serum AlbuminCrosslinkingImplant CoatingsSurface Coatings For Implants
Titanium surfaces were coated with bovine serum albumin using carbodiimide, a crosslinking agent. The durability of the coated surfaces and the inhibitory effect of the albumin coating on bacterial adherence were tested in vitro for 20 consecutive days at 37 degrees C in phosphate buffered saline, with intermittent agitation. The results showed that only 10% of the coated bovine serum albumin decayed off the surface during the 20-day incubation period. The inhibition rate of the albumin coating on bacterial adherence remained high (greater than 8.5%) throughout the experiment. The results suggested potential use of this crosslinked albumin coating to reduce bacterial adherence and thus the subsequent possibility of prosthetic or implant infection in vivo.
23