Applied Optics · 2003 · 328 citations · 12 references
Natural SciencesBiophysicsCell BiologyBiophotonicsDigital ImagingOpticsFluorescence MicroscopyNonlinear OpticsOptical ImagingOptical MetrologyPolarization OpticsPolarization-based ImagingBirefringenceBirefringence ImagingData Processing AlgorithmsExtinctionImage Processing AlgorithmsPolarization MicroscopyPolarization-sensitive ImagingPolarized Light Microscopy
We propose image processing algorithms for measuring two-dimensional distributions of linear birefringence using a pair of variable retarders. Several algorithms that use between two and five recorded frames allow us to optimize measurements for speed, sensitivity, and accuracy. We show images of asters, which consist of radial arrays of microtubule polymers recorded with a polarized light microscope equipped with a universal compensator. Our experimental results confirm our theoretical expectations. The lowest noise level of 0.036 nm was obtained when we used the five-frame technique and four-frame algorithm without extinction setting. The two-frame technique allows us to increase the speed of measurement with acceptable image quality.
12
Recent developments in instrumentation in ellipsometry
P.S. Hauge · Surface Science · 1980 · 240 citations
Mueller matrix imaging polarimetry
J. Larry Pezzaniti · Optical Engineering · 1995 · 189 citations
Birefringence of Single and Bundled Microtubules
Rudolf Oldenbourg, Edward D. Salmon, Phong T. Tran · Biophysical Journal · 1998 · 151 citations · Full text