Publication | Closed Access
A TL/OSL Emission Spectrometer Extension of the Ris Reader
15
Citations
6
References
2002
Year
Luminescence SpectraEngineeringRisø ReaderRis ReaderOptical DiagnosticsSpectroscopyOptical PropertiesMass SpectrometryLuminescence LightUv-vis SpectroscopyAtomic Emission SpectroscopyFiber OpticsInstrumentationOptical CharacterizationLuminescence PropertyOptical SensorSpectrochemical AnalysisOptical Sensors
The integration of a low cost, compact sized spectrometer with the Risø reader is described. The luminescence light emitted by the sample is transmitted by an optical fibre onto a fixed entrance slit of a spectrograph. The light is measured with a high sensitivity 2048-element CCD-linear array detector (Avantes PC2000) sensitive in the UV-VIS region. The CCD array has a low readout noise and a photon sensitivity of 86 photons per count. Examples are given of both thermally and optically stimulated 3-D spectra showing the luminescence spectra as a function of temperature and time respectively. Spectra from CaF2:Tm (TLD-300) chips could be distinguished from the background (2 SD) at a 0.7 Gy dose level.