Magnetic Resonance in Medicine · 2004 · 33 citations · 34 references
EngineeringBioengineeringBiophysicsBiomedical ImagingBiosensorsMagnetic Resonance BiophysicsMagnetic Resonance SpectroscopyReactive Oxygen SpeciesBlack CarbonBlood FlowCarbon BlackCarbon MaterialsOxygen SensingOxygen SensorsTissue OximetryTissue Oxygenation Measurement
New electron paramagnetic resonance (EPR) oximetry probes were identified in the class of carbon black materials. These compounds exhibit very high oxygen sensitivity and favorable EPR characteristics for biological applications. At low pO(2), the linewidth is particularly sensitive to changes in oxygen tension (sensitivity of 750 mG/mmHg). The application of the probes for oximetry was demonstrated in vivo: the pO(2) was measured in muscle in which the blood flow was temporarily restricted as well as in tumor-bearing mice during a carbogen breathing challenge. The responsiveness to pO(2) was stable in muscle for at least 3 months. No toxicity was observed using these materials in cellular experiments and in histological studies performed 2, 7, and 28 days after implantation. In view of their EPR characteristics (high sensitivity) as well as the well-characterized production procedure that make them available on a large scale, these probes can be considered as very promising tools for future developments in EPR oximetry.
34
Shotaro Ogawa, Ravi S. Menon, David W. Tank et al. · Biophysical Journal · 1993 · 1.8K citations · Full text
Guanglong He, Ravi Shankar, Michael Chzhan et al. · Proceedings of the National Academy of Sciences · 1999 · 381 citations