IEEE Transactions on Instrumentation and Measurement · 2010 · 64 citations · 4 references
EngineeringElectrical EngineeringMechatronics EngineeringInstrumentationOpticsOptoelectronicsAngular MeasurementCalibrationEquipment CalibrationOptical InstrumentationOptical MetrologySensor CalibrationAngle Measurement Error SourcesAngular Measurement SystemsAngular MetrologyCalibration Of Optical InstrumentsGyroscope CalibrationInstrumentation CalibrationLaser Metrology SystemsReal-time Processing
This paper deals with the installation errors of an optical angular encoder (OAE). A harmonic model is proposed to describe the installation errors, including the effects of eccentricity, swash, and distortion. Moreover, a method of OAE calibration with the ring laser gyroscope (RLG) is introduced. The calibration method can potentially reach the accuracy of 0.1". The calibration method is proved to be experimentally effective. With the harmonic model obtained from the experiment, the calibration of the OAE can be carried out in real time.
4
Julio Gea-Banacloche, Leno M. Pedrotti, Virgil E. Sanders et al. · Reviews of Modern Physics · 1985 · 699 citations
Leonel N. Menegozzi, Willis E. Lamb · Physical review. A, General physics · 1973 · 136 citations
Natural Sciences, Physics, Optics +16