Applied Optics · 1992 · 45 citations · 10 references
EngineeringBiomedical EngineeringDiagnostic ImagingDigital ImagingRadiation DetectionDetector TechnologyDigital Image ProcessingImage ProcessingOptical Transfer FunctionPhotodetectorsRadiation Imaging DetectorsDetector CharacterizationDetectorsImaging DetectorsModulation Transfer FunctionModulation Transfer Function AnalysisOptical Transfer Function MeasurementPixel Detectors
A modulation transfer function (MTF) measurement technique suitable for large-format, small-pixel detector characterization has been investigated. A volume interference grating is used as a test image instead of the bar or sine wave target images normally used. This technique permits a high-contrast, large-area, sinusoidal intensity distribution to illuminate the device being tested, avoiding the need to deconvolve raw data with imaging system characteristics. A high-confidence MTF result at spatial frequencies near 200 cycles/mm is obtained. We present results at several visible light wavelengths with a 6.8-microm-pixel CCD. Pixel response functions are derived from the MTF results.
10
Scientific Charge-Coupled Devices
James R. Janesick, Tom Elliott, S. A. Collins et al. · Optical Engineering · 1987 · 862 citations
Engineering, Physics, Photonics +16