Measurement of lateral charge diffusion in thick, fully depleted, back-illuminated CCDs

Armin Karcher, C. Bebek, William F. Kolbe, Dominic Maurath, Vignesh Prasad, M. Uslenghi, Martin Wagner

IEEE Transactions on Nuclear Science · 2004 · 38 citations · 3 references

DOIFull text

Open access

Concepts

Abstract

Lateral charge diffusion in back-illuminated CCDs directly affects the point spread function (PSF) and spatial resolution of an imaging device. This can be of particular concern in thick, back-illuminated CCDs. We describe a technique of measuring this diffusion and present PSF measurements for an 800/spl times/1100, 15 /spl mu/m pixel, 280 /spl mu/m thick, back-illuminated, p-channel CCD that can be over-depleted. The PSF is measured over a wavelength range of 450 nm to 650 nm and at substrate bias voltages between 6 V and 80 V.

References

3