Publication | Closed Access
Automatic RF voltage calibration with a primary voltage standard up to 1 GHz
10
Citations
3
References
1993
Year
Sensor CalibrationElectrical EngineeringEngineeringAutomatic SystemRadio FrequencyMeasurementCalibrationComputer EngineeringPrimary VoltageEducationMicrowave MeasurementThermal Voltage ConverterInstrumentationPrimary Voltage StandardMicrowave EngineeringRf SubsystemHigh-frequency MeasurementElectromagnetic Compatibility
An automatic system for the calibration of RF voltage transfer standards in the frequency range between 1 MHz and 1 GHz is described. It incorporates a coaxial dual dry-load calorimeter as a primary voltage standard, with an automated computer-controlled DC voltage balance. By means of comparison measurements with a thermal voltage converter, the estimated relative uncertainties (2 sigma ) for frequencies between 1 MHz and 100 MHz have been reduced by a factor of two to values between 0.01% and 0.1%.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
| Year | Citations | |
|---|---|---|
Page 1
Page 1