Microscale Thermophysical Engineering · 1998 · 51 citations · 12 references
EngineeringBiomedical EngineeringAnalytical ChemistryInstrumentationMetrologyOptical MetrologyOptical SensorsRefractive IndexConcentrationHigh-precision MeasurementLaser DiagnosticsLaser-based DiagnosticsRefractive Index DetectionRefractive Index DeterminationRefractive Index MeasurementRefractive Index Sensing
Measurement of liquid concentration is important in fields such as chemical analysis and processing, diagnostics, and semiconductor manufacturing. This work presents a highprecision, noncontact, laser-based technique to measure the refractive index of liquid mixtures and solutions, which can be used to determine the concentration of a mixture. The technique can resolve extremely small changes in concentration, on the order of 0.01%. The essential components in the system include a HeNe laser as the light source, a precision glass cuvette as the liquid container and optical cell, and a very-high-resolution position sensor and high-precision digital voltmeter. The experimental configuration is small, reliable, and inexpensive. Results are presented for ethanol - water and methanol - water mixtures and two water - salt solutions, with very good agreement found between measured values and those reported in the literature.
12
Optical Constants of Water in the 200-nm to 200-μm Wavelength Region
George M. Hale, Marvin R. Querry · Applied Optics · 1973 · 4.4K citations
Concentration measurements in a transverse jet by planar laser-induced fluorescence of acetone
Antonio Lozano, S. H. Smith, M. G. Mungal et al. · AIAA Journal · 1994 · 77 citations
Natural Sciences, Chemistry, Physics +16