Sensors · 2003 · 33 citations · 15 references
EngineeringBioengineeringChemical EngineeringBiosensorsElectrochemistrySensorsBiological SensorsBiosensor DesignElectrochemical BiosensorsElectrochemical SensorsEnzymatic BiosensorsEnzyme-based BiosensorsBiosensor SelectivityCapacitanceOrganophosphate BiosensorsOrganophosphate CompoundsOrganophosphate PesticidesOrganophosphorus PesticidesPesticide DetectionPotentiometric Enzyme Biosensors
The realisation of a miniaturised potentiometric enzyme biosensor is presented. The biosensor chip utilises the enzyme organophosphorus hydrolase (OPH) for the direct determination of pesticides. The transducer structure of the sensors chip consists of a pH-sensitive capacitive electrolyte-insulator-semiconductor (EIS) structure that reacts towards pH changes caused by the OPH-catalised hydrolysis of the organophosphate compounds. The biosensor is operated versus a conventional Ag/AgCl reference electrode. Measurements were performed in the capacitance/voltage (C/V) and the constant capacitance (ConCap) mode for the two different pesticides paraoxon and parathion. For the development of this new type of biosensor, different immobilisation strategies, influence of buffer composition and concentration, transducer material, detection limit, long-term stability and selectivity have been studied.
15
Anup K. Singh, A.W Flounders, Joanne V. Volponi et al. · Biosensors and Bioelectronics · 1999 · 154 citations
New Trends in Biosensors for Organophosphorus Pesticides