Applied Optics · 2011 · 73 citations · 21 references
Natural SciencesPhysicsOptical MaterialsOpticsSpectroscopyAtmospheric Refraction CorrectionGas SensorsLaser SpectroscopyOptical MetrologyOptical SpectroscopyRefractive IndexCoverage FactorFabry-perot ResonatorsIndex Of Refraction MeasurementRefractive Index DeterminationRefractive Index MeasurementRefractometryTemperature Measurement Systems
We present a method of measuring the refractive index of dry gases absolutely at 632.8 nm wavelength using a Fabry-Perot cavity with an expanded uncertainty of <3×10⁻⁹ (coverage factor k=2). The main contribution to this uncertainty is how well vacuum-to-atmosphere compression effects (physical length variation) in the cavities can be corrected. This paper describes the technique and reports reference values for the refractive indices of nitrogen and argon gases at 100 kPa and 20 °C with an expanded uncertainty of <9×10⁻⁹ (coverage factor k=2), with the additional and larger part of this uncertainty coming from the pressure and temperature measurement.
21
Refractive index of air: new equations for the visible and near infrared
P.E. Ciddor · Applied Optics · 1996 · 1.4K citations
Bengt Edlén · Metrologia · 1966 · 1.2K citations
An Updated Edlén Equation for the Refractive Index of Air
K P Birch, M J Downs · Metrologia · 1993 · 596 citations
High-accuracy displacement interferometry refin air
William T. Estler · Applied Optics · 1985 · 108 citations