IEEE Transactions on Biomedical Engineering · 2010 · 82 citations · 13 references
Veterinary MedicineAnesthesiologyAnesthesiaIntravenous SedationProcess ControlAnesthetic MonitoringMonitoring During SedationPatient MonitoringPerioperative MonitoringClosed-loop StabilityCross-correlation AnalysisDelay CompensationDelay EstimationSmith PredictorTime Delay EstimationTime-varying DelaysVariable Time Delays
The presence of artifacts plays a crucial role in automatic sedation systems and may introduce variable time delays (TDs) in the closed-loop-control structures. This paper presents a successful procedure to estimate the varying TD of the bispectral index (BIS) monitor used in closed-loop control during intensive care. The TD estimation (TDE) is based on the cross-correlation analysis technique and the method is validated with real measured signals of propofol and BIS. Extended prediction self-adaptive control is used in combination with a Smith predictor to reduce the computational burden imposed by the variable TD. The conclusion is that an online TDE of the BIS monitor improves the performance of the closed-loop system for reference tracking, disturbance rejection, and overall stability.
13
System identification—Theory for the user
Naresh K. Sinha · Automatica · 1989 · 9.2K citations
Michel Struys, Hugo Vereecke, Anneliese Moerman et al. · Anesthesiology · 2003 · 221 citations · Full text
Time Delay of Index Calculation
Stefanie Pilge, Robert Zanner, Gerhard Schneider et al. · Anesthesiology · 2006 · 186 citations · Full text
Medicine, Anesthesiology, Anesthesia +15