Surface and Interface Analysis · 1990 · 101 citations · 59 references
Natural SciencesChemistryAnalytical ChemistryElemental AnalysisSpectroscopyAtomic Emission SpectroscopyAtomic SpectroscopyElectronic SpectroscopyEmission SpectroscopyPhotoelectron SpectroscopyX-ray Photoelectron SpectroscopyAtomic Emission SpectrometryCalibration In Atomic SpectroscopyDetector Sensitivity CalibrationEmission SpectraEmission SpectrumEmission Spectrum CalculationEnergy DependenceTheoretical Terms
The energy dependence of spectrometer transmission and detector sensitivity must be known to use published sensitivity factors in quantitative AES and XPS, and consistent instrument results, theoretical quantification, and reference data banks rely on this knowledge. We develop simple experimental procedures to determine these dependencies. We performed detailed measurements on a modified VG Scientific ESCALAB II metrology spectrometer in both constant ΔE/E and constant ΔE modes and compared the results with theoretical estimates. The experimental procedures outperform detailed electron‑optical calculations, allowing removal of transmission and sensitivity terms to convert measured spectra into true electron‑emission spectra and establishing a first step toward reference‑sample calibration for all instruments.
Abstract For the use of published general or theoretical sensitivity factors in quantitative AES and XPS, the energy dependence of both the spectrometer transmission function and the detector sensitivity must be known. Here, we develop simple procedures that allow these dependencies to be determined experimentally. Detailed measurements for a modified VG Scientific ESCALAB II, the metrology spectrometer , operated in both the constant Δ E / E and constant Δ E modes, are presented and compared with theoretical estimates. It is shown that an exceptionally detailed electron‐optical calculation, involving proprietary information, would be required to match the accuracy of the experimental procedures developed. Removal of the spectrometer transmission function and the detector sensitivity terms allow the measured spectrum to be converted to the true electron emission spectrum, irrespective of the mode of operation. This provides the first step to the provision of reference samples to calibrate the transmission functions and detector sensitivities of all instruments so that they, in turn, may produce true electron emission spectra. This is vital if: (1) all instruments are to give consistent results; (2) theoretical terms are to be used in quantifying either AES or XPS; (3) reference data banks are to be established for AES or XPS.
59
Handbook of X-Ray Photoelectron Spectroscopy
J. F. Moulder · Medical Entomology and Zoology · 1995 · 23.7K citations