Measured X-Ray Spectra from 25 to 110 kVp for a Typical Diagnostic Unit

Robert G. Waggener, Louis B. Levy, Lee F. Rogers, Peter Zanca

Radiology · 1972 · 22 citations · 6 references

Concepts

Abstract

Diagnostic x-ray spectra were measured with a NaI(T1) spectrometer system. A digital computer was used to correct the observed spectrum and to find the true spectrum by solving a matrix equation. X-ray spectra for various diagnostic x-ray units were measured under different conditions of focal spot size, duration of exposure, tube current, and kVp. Half-value layer, effective photon energy, and mean energy were measured and compared for similar x-ray units. The principal variation in x-ray spectra was related to differences in kVp and tube filtration. Focal spot size and mAs variation did not significantly influence the measured x-ray spectra.

References

6