Rapid Specific and Label-Free Detection of Porcine Rotavirus Using Photonic Crystal Biosensors

Leo Li‐Ying Chan, Theresa B. Kuhlenschmidt, Charles J. Choi, Mark S. Kuhlenschmidt, Brian T. Cunningham

IEEE Sensors Journal · 2009 · 80 citations · 28 references

Concepts

Abstract

A rapid, label-free assay using photonic crystal biosensors for the direct detection of intact viable rotaviruses is demonstrated. This paper describes the design and operation of the biosensor, surface functionalization with anti-rotavirus antibody, characterization of the detection sensitivity, and demonstration of specificity with respect to non-target virus particles. A 30-min assay of a partially processed water sample yielded a detection sensitivity of 36 virus focus forming units without the use of any external reagents. This sensitivity is comparable to commercial enzyme-linked immunosorbent assays. These results suggest photonic crystal biosensors may be useful for real-time monitoring of virus contamination of environmental water resources.

References

28