Continuum x-ray source as a calibration system for charge coupled devices

K. Hashimotodani, T. Toneri, Shunji Kitamoto, H. Tsunemi, Kentaro Hayashida, Emi Miyata, K. Katayama, T. Kohomura, R. Asakura, K. Koyama,

Review of Scientific Instruments · 1998 · 12 citations · 5 references

Concepts

Abstract

We present x-ray spectra obtained by a charge coupled device (CCD) from an electron impacting x-ray source using several kinds of electron targets: Au, Al–bronze, Al, and pure Al. The x rays from the source are dispersed by a grating and the dispersed x rays are focused on the CCD. Owing to the fine spatial and moderate energy resolutions of the CCD, fine spectra with 2 eV energy resolution up to 2.2 keV are obtained. X rays from the pure Al target (99.999% purity) provide good continuum x rays except for Al K x rays. This continuum source is useful as a calibration source for low energy x-ray detectors.