IEEE Transactions on Electron Devices · 1993 · 19 citations · 3 references
EngineeringChemistryAnalytical ChemistryInstrumentationSensorsCharged Particle DetectionDetector SystemsIonization ProcessesMass Spectrometric ImagingMass SpectrometryElectron Emission EffectsPhotodiode CircuitsSecondary Electron EmissionSecondary Ion Mass SpectrometrySecondary Ion Mass Spectrometry Imaging
An area detector for a secondary ion mass spectrometry (SIMS) ion microscope and its performance are described. The operational principle is based on detecting the change in potential of a floating photodiode caused by the ion-induced secondary-electron emission and the incoming ion itself. The experiments demonstrated that 10/sup 1/-10/sup 5/ aluminum ions per pixel can be detected with good linear response. Moreover, relative ion sensitivities from hydrogen to lead were constant within a factor of 2. The performance of this area detector provides the potential for detection of kiloelectronvolt ion images with current ion microscopy.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
3
Ion-induced electron emission from clean metals